Remove weight from the pan

Remove weight from the pan. Make
sure the machine is reading zero
before weighing. Do not over load
the pan. If the problem recurs
calibrate. Possible damage to the
internal weighing cell.
OUt2 Out of Zero reading Make sure the pan is fitted correctly.
Make sure nothing is touching the
pan when turning on the unit.
Calibrate the machine. Possible
damage to the internal weighing cell.
UnSt Unstable Make sure the balance is on a flat
surface and away from vibration.
Possible damage to the internal
weighing cell.
LO or Low Battery indicator Change the batteries or use on the
power adapter.
ErrE E-Prom error Software error turn off and restart.
Contact Adam Equipment or your
dealer for further assistance.
No Power when turning on Change the batteries or use on the
power adapter.